TOPwave Installation and Setup
25
Parameter Description Factory
settings
Value
min.
Value
max.
T MIN SET Min. allowed temp. in programs 50°C 0°C 300°C
P MAX SET Max. allowed pressure in programs 130 bar 0 bar 130 bar
P MIN SET Min. allowed pressure in programs 0 bar 0 bar 130 bar
POWER MAX Max. allowed power in programs 100 % 0 % 100 %
POWER MIN Min. allowed power in programs 0 % 0 % 100 %
STANDARD APPLI-
CATION
Application, which is to be active af-
ter the device is switched on
1 1 198
USER MANAGE-
MENT ON
/OFF
If the U
SER MANAGEMENT is activated
(ON), as a first screen by switching
on the control unit, the user login
appears. Is
OFF displayed the Main
menu will be shown.
off off on
PRESSURE ON/OFF The OPC pressure measurement
can generally be activated or deacti-
vated
off off on
MAX.
DIGESTION TIME
Max. runtime of a temperature pro-
gram
120 min 0 min 120 min
T MIN SET Min. allowed temp. in programs 50°C 0°C 300°C
WEB VISU Activates web visualization 0 (off) 0 (off) 1 (on)
MASTER CODE Login-Code of master
enter 4-digit code
0 0 9999
USER CODE Login-Code of user
enter 4-digit code
0 0 9999
FILE WARNING Warning is shown if there are more
than 100 files
(reasonable value: 100)
100 1 1000
COUNT U/D (1/0) Total runtime can be adjusted as a
count-up or countdown
1 0
(down)
1 (up)
Open with [
NEXT] the 2
nd
SETUP menu (parameters in table below).
Select the language by pressing a language button.
Select date and time by pressing [
SET TIME].