TOPwave Operating
41
1 Coupling cap
2 Rupture disc
3 Lid
4 Vessel bottom
Fit the rotor symmetrically
(see section "Positioning of the vessels in the rotor" p. 77).
Screw the vent fitting onto the associated collection vessel tube (it may be necessary
to first move the vessel into the proper position).
1 Vent fitting
2 Vessel tube
Seal any unused positions on the collection vessel with the supplied blind stoppers or
with lock screws in order to ensure the proper operation of the gas removal.
Blind stopper
Mounting vessels into the
rotor