Care TOPwave
64
If the foil has strong signs of wear (flaking, blistering, discoloration) you have to renew
the polarization foil.
Remove the old foil from the glass prism.
Clean the glass prism thoroughly. You can remove old sticking remains with an al-
coholic solution.
Remove the protecting foil from the self-adhesive side of the polarization foil.
Stick the polarization foil onto the polished longer side of the glass prism and press
the foil on.
1 Polarization foil
2 Glass prism
Only then remove the protection foil from the over side of the polarization foil.
Cut off remains with a sharp knife or scissors.
Insert the glass prism into the copling cap und fix it (see above).
For assembly and dismantling of the CX 100 pressure control cap the same procedure
must be followed how descript above. Additionally, a cover plate is clipped onto the top
of the cap.
For cleaning the side of the cover plate is raised using a pincette or screwdriver and
removed.
If the cover plate has bulged upwards as a result of safety venting it can be reused by
simple rotating it.
1 Cover plate
2 Safety venting
Renewing the polarization
foil
Cap for pressure control of
CX 100