Specification TOPwave
72
Vessel type Volu-
mes
max.
Pressure
max. Tempe-
rature
(continuous
operation)
max. sample
weight
(organic)
Min. fill
volume (a-
cid)
PM 40 40 ml 40 bar
(580 psi)
210 °C **
(410 °F)
< 300 mg > 7 ml
PM 60 60 ml 40 bar
#
(580 psi)
210 °C **
(410 °F)
< 500 mg > 7 ml
PL 100 100 ml 40 bar *
(580 psi)
210 °C **
(410 °F)
< 500 mg > 7 ml
CX 100 100 ml 100 bar
(1450 psi)
250 °C
(482 °F)
< 1000 mg > 7ml
QX 20/
QX 22
20 ml 100 bar
(1450 psi)
230 °C
(446 °F)
< 300 mg > 7 ml
Material: TFM-PTFE, quartz, aluminum oxide
*25 bar above 200 °C
# 50 bar with special rupture discs (not enclosed in standard delivery)
**Short - term (max. 15 min.) heating to 260 °C (500 °F) is possible with all vessels.
Overpressure protection: Rupture disc for all pressure vessels.