EasyManua.ls Logo

Endress+Hauser Deltabar S FMD76 Technical Information

Endress+Hauser Deltabar S FMD76
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
TI382P/00/EN/06.09
No. 71095431
Technical Information
Deltabar S PMD70/75, FMD76/77/78
Differential pressure measurement
Differential pressure transmitter with ceramic and silicon sensors
Overload-resistant and function-monitored, Communication via
HART, PROFIBUS PA or FOUNDATION Fieldbus
Application
The Deltabar S differential pressure transmitter is used
for the following measuring tasks:
Flow measurement (volume or mass flow) in
conjunction with primary elements in gases, vapours
and liquids
Level, volume or mass measurement in liquids
Differential pressure monitoring, e.g. of filters and
pumps
International usage thanks to a wide range of approvals
Your benefits
Very good reproducibility and long-term stability
High reference accuracy: up to 0.075%,
as PLATINUM version: 0.05%
Turn down 100:1, higher on request
Used for flow and differential pressure monitoring
up to SIL 3, certified to IEC 61508 by TÜV SÜD
HistoROM
®
/M-DAT memory module
Function-monitored from the measuring cell to the
electronics
Continuous modularity for differential pressure,
hydrostatic and pressure (Deltabar S, Deltapilot S,
Cerabar S), e.g.
replaceable display
universal electronics for pressure and
differential pressure
Quick commissioning thanks to quick setup menu
Easy and safe menu-guided operation on-site,
via 4...20 mA with HART, via PROFIBUS PA or
via FOUNDATION Fieldbus
Extensive diagnostic functions

Table of Contents

Other manuals for Endress+Hauser Deltabar S FMD76

Questions and Answers:

Endress+Hauser Deltabar S FMD76 Specifications

General IconGeneral
Device TypeDifferential Pressure Transmitter
Process Pressure Limitup to 400 bar (5800 psi)
CommunicationHART, PROFIBUS PA, FOUNDATION Fieldbus
Housing MaterialAluminum, stainless steel
ApplicationLevel, volume or mass measurement in liquids
Measuring Rangeup to 40 bar (580 psi)
Process Temperature-40 to 120 °C (-40 to 248 °F)
Accuracy±0.075%
Output Signal4...20 mA HART
Power Supply10.5 to 45 V DC
Sensor TypePiezoresistive
Long-term stability±0.1% of URL/year
Reference accuracy±0.075% of span
Process ConnectionThreaded and flange connections
Protection ClassIP66
Material316L