Deltabar FMD71, FMD72
Level measurement with electronic differential pressure
Electronic differential pressure transmitter with ceramic
and metal sensors
BA01044P/00/EN/06.17
71376028
Valid as of version
01.00.00

Table of Contents

Other manuals for Endress+Hauser Deltabar FMD71

Questions and Answers

  • D
    Danielle BeckSep 9, 2025
    What does error code M431 mean on Endress+Hauser Deltabar FMD71?
    • D
      davenportkimberlySep 9, 2025
      The error code M431 on your Endress+Hauser Measuring Instruments indicates that the calibration carried out would cause the nominal sensor module range to be exceeded or undershot. To address this, check the measuring range, position adjustment, and the setting.
  • L
    Lisa LarsonAug 25, 2025
    Why is there no display on my Endress+Hauser Measuring Instruments?
    • R
      Ryan MckayAug 25, 2025
      If there is no display on your Endress+Hauser Measuring Instruments, it could be due to several reasons. First, the onsite display might be set too bright or too dark. To resolve this, try setting the local display brighter by simultaneously pressing the indicated buttons or darker by pressing the other indicated buttons. Another potential cause is a connector for the local display that is not properly connected, make sure the connector is properly connected. Finally, the onsite display could be defective, in which case it needs to be replaced.
  • M
    martinjamesAug 25, 2025
    What to do if Endress+Hauser Measuring Instruments device measures incorrectly?
    • J
      Jillian DiazAug 26, 2025
      If your Endress+Hauser Measuring Instruments device is measuring incorrectly, the most likely cause is a configuration error. To fix this, check and correct the parameter configuration.
  • E
    Edward ReyesSep 22, 2025
    What does error F002 mean on Endress+Hauser Measuring Instruments?
    • D
      Dustin GoodmanSep 22, 2025
      Error F002 on your Endress+Hauser Measuring Instruments means the sensor module does not suit the device. Contact Endress+Hauser Service.
  • W
    Walter ClarkSep 9, 2025
    What does error code M434 mean on Endress+Hauser Measuring Instruments and how to troubleshoot it?
    • D
      dcaldwellSep 10, 2025
      The error code M434 on your Endress+Hauser Measuring Instruments can be caused by values for calibration being too close together, lower/upper range values undershooting/overshooting sensor module range limits, a replaced sensor module with an unsuitable customer-specific configuration, or an unsuitable download carried out. To troubleshoot this, check the measuring range and the setting.
  • P
    Patrick GonzalezSep 14, 2025
    What does error code F512 mean on Endress+Hauser Deltabar FMD71 Measuring Instruments and how can I resolve it?
    • L
      Logan LloydSep 14, 2025
      The error code F512 on your Endress+Hauser Measuring Instruments indicates that the table is not monotonic. To resolve this, correct and accept the table.
  • W
    William AndersonSep 4, 2025
    What does error code F841 mean on Endress+Hauser Deltabar FMD71 and how to solve it?
    • Y
      ygarciaSep 4, 2025
      The error code F841 on your Endress+Hauser Measuring Instruments indicates overpressure or low pressure, or a defective sensor module. To solve this, check the pressure value.
  • K
    Karen WilsonSep 4, 2025
    What does error code S841 mean on Endress+Hauser Deltabar FMD71 Measuring Instruments?
    • M
      Michael SmithSep 4, 2025
      The error code S841 on your Endress+Hauser Measuring Instruments indicates overpressure or low pressure, or a defective sensor module. To address this, check the pressure value.
  • C
    challSep 23, 2025
    What does error F083 mean on Endress+Hauser Measuring Instruments?
    • B
      bullockjohnathanSep 23, 2025
      Error F083 on your Endress+Hauser Measuring Instruments indicates that electromagnetic effects are outside the permitted range. In this case, contact Endress+Hauser Service or restart the device.
  • C
    Chase GuzmanAug 29, 2025
    What does error code F083 mean on Endress+Hauser Measuring Instruments and what are possible solutions?
    • B
      Briana SanchezAug 30, 2025
      The error code F083 on your Endress+Hauser Measuring Instruments can occur due to electromagnetic effects outside the permitted range or a defective sensor module. You can try restarting the device.

Endress+Hauser Deltabar FMD71 Specifications

General IconGeneral
Device TypeDifferential Pressure Transmitter
Measuring PrinciplePiezoresistive
Output Signal4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus
Power Supply10.5 to 45 V DC
Accuracy±0.065% of span
Process Temperature-40 to 150°C (-40 to 302°F)
Ambient Temperature-40 to 85°C (-40 to 185°F)
Process ConnectionThread, flange, hygienic connections
Housing MaterialAluminum
Diaphragm Material316L Stainless Steel
CommunicationHART, PROFIBUS PA, FOUNDATION Fieldbus
ApprovalsATEX, IECEx
Protection ClassIP67
Process Pressure (Absolute)Up to 40 bar
ApplicationLevel, flow, and differential pressure measurement in liquids, gases, and vapors

Related product manuals