Deltabar S FOUNDATION Fieldbus Installation
Endress+Hauser 19
The standard mounting bracket version is not suitable for use in an application subject to
vibrations.
The vibration resistance of the reinforced version of the mounting bracket has been tested
according to IEC 61298-3, see the "Vibration resistance" section in the technical documentation
TI00382P.
When using a valve block, the block's dimensions must be taken into account.
Bracket for wall and pipe mounting including retaining bracket for pipe mounting and two nuts.
material of the screws used to secure the device depend on the order code.
Technical data (e.g. dimensions or order numbers for screws) see accessory document
SD01553P/00/EN.
Incorrect handling!
Damage of the device!
‣ Disassembly of the screws with item number (1) is not permissible under any circumstances
and will result in loss of warranty.
Standard design Heavy duty design
A0031326 A0031327