Ultrasonic Level Measurement
prosonic M FMU 40/41/43
Compact transmitters for non-contact level
measurement of fluids, pastes and coarse
bulk materials
Application
The compact Prosonic M transmitters are
used for continuous, non-contact level
measurement in fluids and coarse bulk
materials.
Additionally, the sensors can be used for
flow measurement in open channels and
measuring weirs.
The following interfaces are available for
system integration:
HART
®
(standard), 4 to 20 mA
PROFIBUS-PA
Foundation Fieldbus
The maximum measuring range for
Prosonic M sensors are:
FMU 40:
Fluids, 16 feet (5 m)
Bulk solids, 6 feet (2 m)
FMU 41:
Fluids, 26 feet (8 m)
Bulk solids, 12 feet (3.5 m)
FMU 43:
Fluids, 50 feet (15 m)
Bulk solids, 23 feet (7 m)
Features and benefits
Simple, menu-guided on-site
operation with four-line plain text
display
Envelope curves on the on-site
display for simple diagnosis
Easy operation, diagnosis and
measuring point documentation with
the supplied ToF Tool operating
program
Alignable NEMA 6P (IP 68) aluminum
housing
Optional remote display and operation
Installation via 1-1/2” NPT, 2” NPT
or 4” universal slip-on flange
Integrated temperature sensor for
Time-of-Flight correction provides
accurate measurements, even with
temperature changes
Linearization function (up to 32 points)
for measured value output in any unit
of length, volume, or flow rate
Non-contact measurement method,
therefore almost independent of
product properties
Technical
Information
TI 365F/24/ae
The Power of Know How
FMU 40
FMU 43
FMU 41
Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser FMU 40 and is the answer not in the manual?

Endress+Hauser FMU 40 Specifications

General IconGeneral
BrandEndress+Hauser
ModelFMU 40
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals