XA00639F-A Levelflex FMP51/52/53/54/55/56/57
Endress+Hauser 9
Electronics compartment Ex i
Service interface (CDI)
Taking the following values into consideration, the device can be connected to the Endress+Hauser FXA291
service tool or a similar interface:
Service interface
U
i
= 7.3 V
effective inner inductance L
i
= negligible
effective inner capacitance C
i
= negligible
U
o
= 7.3 V
I
o
= 100 mA
P
o
= 160 mW
permitted outer inductance L
o
= negligible
permitted outer capacitance C
o
= negligible