Operating Instructions
Levelflex FMP51, FMP52, FMP54
FOUNDATION Fieldbus
Guided wave radar
BA01052F/00/EN/04.16
71344062
01.01.zz (Device firmware)

Table of Contents

Other manuals for Endress+Hauser Levelflex FMP54

Questions and Answers

  • T
    Timothy PhillipsSep 9, 2025
    Why does the measured total level jump to lower values during filling when 'Partially filled' is selected in 'Tank level' on Endress+Hauser Measuring Instruments?
    • M
      Michelle MillerSep 9, 2025
      If, when 'Partially filled' is selected in 'Tank level', the measured total level jumps to lower values during filling operations, the total level may be running into the upper blocking distance. To resolve this, reduce the blocking distance.
  • J
    Jason MasonAug 25, 2025
    How to fix the wrong slope of the measured interface level in Endress+Hauser Measuring Instruments?
    • M
      matthewbushAug 25, 2025
      The wrong slope of the measured interface level is caused by the wrong dielectric constant (DC value). To correct this, enter the correct dielectric constant (DC value) of the upper medium.
  • J
    Jessica GarrettAug 25, 2025
    Why Endress+Hauser Radar device does not respond?
    • J
      Jason CarpenterAug 26, 2025
      If your Endress+Hauser Radar device isn't responding, it could be due to a couple of reasons: * The supply voltage might not be correctly connected; ensure you connect the correct voltage. * The cables might not be making proper contact with the terminals; make sure there's good electrical contact between the cable and the terminal.
  • E
    Elizabeth JenkinsSep 23, 2025
    Why is there a wrong slope of the measured interface level in my Endress+Hauser Levelflex FMP54 Measuring Instruments?
    • C
      Cynthia PattonSep 23, 2025
      If there is a wrong slope of the measured interface level in your Endress+Hauser Measuring Instruments, the dielectric constant (DC value) may be incorrect. Enter the correct dielectric constant (DC value) of the upper medium.
  • S
    Scott MercerSep 10, 2025
    What to do if my Endress+Hauser Measuring Instruments device does not respond?
    • L
      lorirogersSep 10, 2025
      If your Endress+Hauser Measuring Instruments device isn't responding, there are a couple of things to check: * First, make sure the supply voltage is correctly connected. If it is not, connect the correct voltage. * Second, verify that the cables are making proper contact with the terminals. Ensure electrical contact between the cable and the terminal.
  • M
    matakerriSep 14, 2025
    What to do if the values on my Endress+Hauser Measuring Instruments display are invisible?
    • M
      martin13Sep 15, 2025
      If the values on your Endress+Hauser Measuring Instruments display are not visible: * The contrast setting may be too weak or too strong. Increase contrast by pressing the indicated buttons simultaneously or decrease contrast by pressing the indicated buttons simultaneously. * The plug of the display cable might not be connected correctly. Connect the plug correctly. * The display itself could be defective. Replace the display.
  • B
    Barbara MartinezSep 4, 2025
    Why is the output current less than 3.6 mA on my Endress+Hauser Radar?
    • R
      Ryan JacksonSep 4, 2025
      If the output current is less than 3.6 mA on your Endress+Hauser Radar device: * Check the signal cable connection to ensure it is correct. * Replace the electronics, as the electronics might be defective.
  • D
    Dakota JohnsonSep 4, 2025
    Why CDI communication does not work for Endress+Hauser Radar?
    • R
      Robin WilliamsSep 5, 2025
      If CDI communication does not work with your Endress+Hauser Radar, the COM port on the computer might be set incorrectly. Check the setting of the COM port on the computer and change it if necessary.
  • K
    Kim RodriguezSep 23, 2025
    What to do if with the setting Tank level = Partially filled, the measured total level jumps to lower values during filling operations in Endress+Hauser Measuring Instruments?
    • T
      Tracy ShahSep 23, 2025
      If, with the setting Tank level = Partially filled, the measured total level jumps to lower values during filling operations in your Endress+Hauser Measuring Instruments, the total level might be running into the upper blocking distance. Decrease the blocking distance.
  • A
    Abigail MullinsAug 29, 2025
    How to troubleshoot an Endress+Hauser Radar device that does not respond?
    • M
      Michael GeorgeAug 29, 2025
      If the Endress+Hauser Radar device does not respond, it could be due to the following reasons: * The supply voltage is not connected. Connect the correct voltage. * The cables do not contact the terminals properly. Ensure electrical contact between the cable and the terminal.

Summary

Basic Safety Instructions

Requirements for the Personnel

Outlines necessary qualifications for personnel.

Designated Use

Specifies intended application and materials.

Workplace Safety

Details safety measures for working with the device.

Operational Safety

Covers safe operation, conversions, and repairs.

Incoming Acceptance and Product Identification

Mounting

Mounting Requirements

Covers general requirements and suitable positions.

Notes on the Mechanical Load of the Probe

Details tensile and bending strength limits.

Notes on the Process Connection

Discusses threaded connections and seals.

Mounting Cladded Flanges

Details mounting instructions for cladded flanges.

Securing the Probe

Covers securing rope probes.

FMP54 with Gas Phase Compensation: Mounting the Probe Rod

Details probe rod mounting for FMP54 with GPC.

Mounting the Device

Covers device mounting with thread and flange.

Mounting the Sensor Remote Version

Details mounting for remote sensor versions.

Post-Installation Check

Lists final checks after installation.

Electrical Connection

Connection Conditions

Details terminal assignment and cable specifications.

Cable Specification

Provides details on cable types and specifications.

Power Supply

Details power supply requirements for different approvals.

Connecting the Device

Step-by-step guide for connecting the device.

Post-Connection Check

Lists checks after connecting the device.

Operation Options

Integration into a FOUNDATION Fieldbus Network

Commissioning via Wizard

Commissioning via Operating Menu

Commissioning (Block-Based Operation)

Block Configuration

Guides on configuring blocks for operation.

Configuration of a Level Measurement

Step-by-step guide for configuring level measurement.

Configuration of an Interface Measurement

Guide for configuring interface measurements.

Diagnostics and Troubleshooting

General Trouble Shooting

Provides solutions for common device issues.

Calling Up Remedial Measures

Guides on accessing remedial measures for diagnostic events.

Repairs

Operating Menu

Setup Menu

Details parameters within the Setup menu.

Diagnostics Menu

Outlines diagnostic functions and messages.

Endress+Hauser Levelflex FMP54 Specifications

General IconGeneral
Measuring PrincipleTDR (Time Domain Reflectometry)
Measuring RangeUp to 30 m
Output4-20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus
Power Supply10.5 to 36 V DC (two-wire)
Housing MaterialAluminum, Stainless Steel
Process ConnectionThread, Flange
ApprovalsATEX, IECEx
MaterialPTFE
Protection ClassIP66/IP67