Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus Operating menu
Endress+Hauser 235
17.4.6 "Data logging" submenu
Navigation  Diagnostics → Data logging
Assign channel 1 to 4
Navigation  Diagnostics → Data logging → Assign chan. 1 to 4
Description Allocate a process variable to the respective data logging channel.
Selection Off
Level linearized
Distance
Unfiltered distance
Interface linearized
*
Interface distance
*
Unfiltered interface distance
Thickness upper layer
*
Terminal voltage
Electronic temperature
Measured capacitance
*
Absolute echo amplitude
Relative echo amplitude
Absolute interface amplitude
*
Relative interface amplitude
*
Absolute EOP amplitude
EOP shift
Noise of signal
Calculated DC value
*
Analog output adv. diagnostics 1
Analog output adv. diagnostics 2
Analog output 1
Analog output 2
Analog output 3
Analog output 4
Factory setting Off
Additional information A total of 1000 measured values can be logged. This means:
1000 data points if 1 logging channel is used
500 data points if 2 logging channels are used
333 data points if 3 logging channels are used
250 data points if 4 logging channels are used
If the maximum number of data points is reached, the oldest data points in the data log
are cyclically overwritten in such a way that the last 1000, 500, 333 or 250 measured
values are always in the log (ring memory principle).
The logged data are deleted if a new option is selected in this parameter.
* Visibility depends on order options or device settings