Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus Integration into a FOUNDATION Fieldbus network
Endress+Hauser 89
Name Label Index Data type Size
(Bytes)
Storage
Class
Write
access
MODE_BLK Description
decimal_places_menu Decimal places menu 88 ENUM16 2 Static x AUTO
level_unit_ro Level unit 90 ENUM16 2 Static x OOS
operating_mode_ro Operating mode 91 ENUM16 2 Static x OOS
assign_channel_1 Assign channel 1 92 ENUM16 2 Static x AUTO
assign_channel_2 Assign channel 2 93 ENUM16 2 Static x AUTO
assign_channel_3 Assign channel 3 94 ENUM16 2 Static x AUTO
assign_channel_4 Assign channel 4 95 ENUM16 2 Static x AUTO
clear_logging_data Clear logging data 97 ENUM16 2 Static x AUTO
logging_interval Logging interval 96 FLOAT 4 Static x AUTO
display_filter_options Filter options 99 ENUM8 1 Static x AUTO
locking_status Locking status 108 BIT_ENUM16 2 Dynamic
distance_unit_ro Distance unit 89 ENUM16 2 Static x OOS
9.6.5 Expert Configuration Transducer Block
The parameters of the Expert Configuration Transducer Block are described in
GP01015F: "Levelflex FMP5x - Description of Device Parameters - FOUNDATION
Fieldbus"
Name Label Index Data type Size (Bytes) Storage Class Write access MODE_BLK
acknowledge_alarm Acknowledge alarm 81 ENUM16 2 Static x AUTO
integration_time Integration time 67 FLOAT 4 Static x OOS
result_self_check Result self check 77 ENUM16 2 Dynamic
start_self_check Start self check 76 ENUM16 2 Static x AUTO
broken_probe_detection Broken probe detection 75 ENUM16 2 Static x AUTO
gpc_mode GPC mode 68 ENUM16 2 Static x OOS
reference_echo_threshold Reference echo threshold 73 FLOAT 4 Static x OOS
const_gpc_factor Const. GPC factor 74 FLOAT 4 Static x OOS
build_up_ratio Build-up ratio 90 FLOAT 4 Dynamic
build_up_threshold Build-up thres. 91 FLOAT 4 Static x AUTO
delay_time_echo_lost Delay time echo lost 78 FLOAT 4 Static x AUTO
empty_capacity Empty capacity 92 FLOAT 4 Static x AUTO
external_pressure_selector External pressure selector 69 ENUM16 2 Static x OOS
measured_capacity Measured capacitance 89 FLOAT 4 Dynamic
gas_phase_compens_factor Gas phase compensation factor 70 FLOT 4 Static x OOS
in_safety_distance In safety distance 80 ENUM16 2 Static x OOS
ratio_amplitude_interface_level Ratio amplitude interface/level 86 FLOAT 4 Static x OOS
interface_criterion Interface criterion 87 FLOAT 4 Dynamic
control_measurement Measurement 106 ENUM16 2 Static x AUTO
control_measurement Control measurement 105 ENUM16 2 Static x AUTO
filter_dead_time Dead time 66 FLOAT 4 Static x OOS
present_reference_distance Present reference distance 72 FLOAT 4 Dynamic
history_reset History reset 83 ENUM16 2 Static x OOS
safety_distance Safety distance 79 FLOAT 4 Static x OOS