Installation Kit CM44x
16 Endress+Hauser
5.2 Installing the module
5.2.1 Bus termination (module 485 only)
If you want to use the internal resistor of the module for bus termination, you must set the hard-
ware switches ( å 27) before installing the module.
5.2.2 Installation conditions
Find out about the upgrading possibilities of your device.
E.g., use the configurator on the product’s web site: www.products.endress.com/cm442
or .../cm444R or .../CSF48 ...
Notes:
You can use a single fieldbus protocol only. Example: HART is activated on your device. After
the installation of the 485 module, you enter an activation code for PROFIBUS DP. The HART
communication will now be deactivated!
When installing the module to a CM442/442R, all possibly existing current outputs will be
deactivated!
You can use a maximum of 2 additional current outputs with a CM444/444R or CM448/448R
device or with CSFxx or CA80XX devices.