EasyManua.ls Logo

Endress+Hauser Liquiphant FTL51B - Device Support; Weld-in Adapter; Mounting Device; Required Tools

Endress+Hauser Liquiphant FTL51B
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XA01810F-C Liquiphant FTL51B, FTL63
14 Endress+Hauser
Accessory high pressure sliding sleeve
The high pressure sliding sleeve can be used for a continuous setting of
the switch point and is suited for zone separation if mounted properly
(see Operating Instructions).
Intrinsic safety
The device is only suitable for connection to certified, intrinsically
safe equipment with explosion protection Ex ia / Ex ib.
The intrinsically safe input power circuit of the device is isolated from
ground. The dielectric strength is at least 500 V
rms
.
Potential equalization
Integrate the device into the local potential equalization.
Optional specification, ID Px, Rx = PA
Connect the weather protection cover to the local potential
equalization.
Bluetooth® module
Basic specification, Position 3, 4 = A7
If the device is equipped with the Bluetooth® module, no battery is
required or allowed.
Basic specification, Position 3, 4 = A8
If the device is equipped with the Bluetooth® module, a battery is
required.
Removal or replacement of the battery is only permitted in non-
hazardous areas.
Observe the information in the Safety Instructions (XA) included with
the Bluetooth® module.
Safety
instructions:
Zone 0
When using under non-atmospheric pressures and non-atmospheric
temperatures: The sensor part of the device approved for Zone 0 does
not cause any ignition hazards.
Safety
instructions:
Zone separation
Zone 0, Zone 1
Device type FTL51B, FTL63 with Basic specification, Position 9 = A
The zone separation wall of the device is made of stainless steel or high
corrosion-resistant alloy of thickness ≥ 1 mm.
Device type FTL63 with Basic specification, Position 9 = B, D, E
The zone partition wall of the device is made of stainless steel or a
highly corrosion-resistant alloy of thickness from 0.2 to 1 mm.
The probe must not be subjected to abrasive or corrosive medium that
may adversely affect the partition for the zone separation.