Installation Micropilot FMR67B HART
16 Endress+Hauser
Purge air connection
Tool:
Open-ended wrench 13 mm (G 1/4")
Open-ended wrench 14 mm (NPT)
Open-ended wrench 17 mm (NPT "adapter")
Min. torque: 6 Nm (4.4 lbf ft)
Max. torque: 7 Nm
Always use dry purge air
In general, purging should only be performed to the extent necessary as excess purging
can cause mechanical damage (abrasion)
10.3.7 Turning the housing
The housing can be rotated up to 380° by loosening the locking screw.
Your benefits
Easy installation due to optimum alignment of housing
Easily accessible device operation
Optimum readability of the local display (optional)
11 1
B
C D
A
A0046660
A Plastic single compartment housing (no locking screw)
B Aluminum single compartment housing
C Dual compartment housing
D Dual compartment housing, L-form
1 Locking screw
NOTICE
The housing cannot be unscrewed fully.
Loosen the external locking screw by a maximum of 1.5 turns. If the screw is unscrewed
too much or completely (beyond the screw anchor point), small parts (counter disk) can
become loose and fall out.
Tighten the securing screw (hexagon socket 4 mm (0.16 in)) with maximum
3.5 Nm (2.58 lbf ft) ± 0.3 Nm (0.22 lbf ft).