Endress+Hauser 9
4.2 Mounting the measuring device
4.2.1 Horizontal orientation
A
50 (2)
£
7.5 (0.3)
15 (0.59)
50 (2)
£
50 (2)
£
B
-
-
.
A0026303
2 Minimum spacing, horizontal orientation. Unit of measurement mm (in)
A Connection of another device type
B DIN rail in accordance with EN 60715 TH35-7.5/15
A horizontal installation ensures better dissipation of heat than a vertical orientation.