Oxymax COS22 Technical data
Endress+Hauser 31
Temperature-pressure
ratings
p
T
[bar]
[°C]
6
12
135
80
[psi]
174
87
T
[°F]
176
275
-5
23
0
32
A0028771-EN
Minimum flow COS22-*1 (standard sensor): 0.02 m/s (0.07 ft/s)
COS22-*3 (trace sensor): 0.1 m/s (0.33 ft/s)
Chemical resistance Parts in contact with the medium are chemically resistant to:
• Diluted acids and alkalis
• Hot water and superheated steam up to max. 135 ˚C (275 ˚F)
• CO
2
up to 100 %, only with trace sensor COS22-*3
NOTICE
Hydrogen sulfide and ammonia shorten the operating life of the sensor.
‣
Do not use the sensor in applications where it is exposed to hydrogen sulfide or
ammonia vapors.
Cross-sensitivity
COS22-*1/3
Molecular hydrogen causes false low readings and can, in a worst-case scenario, result in
total failure of the sensor.
CIP compatibility Yes (COS22-*1/3)
SIP compatibility Yes, max. 140 °C (284 °F) (COS22-*1/3)
Autoclavability Yes, max. 140 °C (284 °F), max. 30 min. (COS22-*1/3)