5. Strip back the jacket and/or insulation of the wires just enough to
connect to the power terminal block.
6. Connect the green wire (ground) to terminal 4, the brown/white wire
(neutral) to terminal 5 and the black wire (hot) to terminal 6 as
shown in Figure 3–4.
7. Close the power terminal box and latch the heated enclosure door.
Application of conduit lubricant
To ensure proper installation, using STL8 lubricant on all conduit screw thread
and its tapped opening is recommended.
STL8 Screw Thread Lubricant is a lithium based, anti-galling substance with
excellent adhesion that maintains rain-tightness and grounding continuity
between conduit fittings. This lubricant has proven very effective between parts
made of dissimilar metals, and is stable in temperatures from -20
F to +300 F.
Do not use this lubricant on exposed current-carrying parts.
Figure 3–4 AC connection terminal block for 120 VAC
enclosure heater
NEU
TERMINAL 5
LINE
TERMINAL 6
TERMINAL 4