Turbimax CUS51D
10 Endress+Hauser
Installation conditions
Installation instructions Mounting applications:
• with retractable assembly Cleanfit W CUA451
• with wastewater assembly Flexdip CYA112 and holder system Flexdip CYH112
• with flow assembly Flowfit CYA251
a0013009
Installation with retractable assembly
Arrow 1 shows the flow direction.
The installation angle α must not exceed 90°.
The recommended installation angle is 75°.
The optical windows of the sensor have to be aligned parallel to
the flow direction (α = 90°) or face the flow direction
(α < 90°).
For manual insertion/retraction of the assembly the medium
pressure may not exceed 2 bar (29 psi).
a0013010
Installation with wastewater assembly
The arrow shows the flow direction.
The installation angle is 45° (recommended) or 90°.
If you use the sensor in open basins, install the sensor in a way
no bubbles can build up around the optical windows.
If you use the sensor in strong aerated basins install the sensor
in an installation angle of 90° to minimize the influence of
bubbles.
a0015169
Installation with flow assembly CYA251
The installation angle is 90°.