Centrifuge 5804/5804 R and 5810/5810 R
10 Descriptions
10-16 Descriptions
Setting the serial address of the centrifuge
The Service programs run completely separately from the normal operating procedure. By simultane-
ously pressing both register keys and by switching on the main power supply at the same time, the user
arrives at the program selection menu.
START Program 1 (SE 1)
Display:
xxxxx = RS232 To set, press the PLUS key,
BBUS the last value set is stored
disable
The device address is set by pressing the START key.
Display:
xx = Device address 20...29 (Hex)
When the PLUS key is pressed, the address which is to be set is incremented.
START – To confirm the address selected
END Program end Press the STOP key.
After the address has been set, the centrifuge must be switched off and switched on again in order to
restore the normal operating status.
Rotor table
The rotor number should be set after the rotor has been exchanged to enable every nominal rotational
speed value to be specified:
1•xxxxx
2•xx
Rotor no. Max. rotational speed Max. radius
2
3
4
5
8
13
14
17
18
A-2-DWP
A-2-MTP
A-4-62
A-4-81
A-4-44
F45-48-PCR
F34-6-38
F45-30-11
T-60-11
3,700 rpm
4,000 rpm
4,000 rpm
4,000 rpm
5,000 rpm
12,000 rpm
12,000 rpm
14,000 rpm
14,000 rpm
147 mm
126 mm
180 mm
180 mm
161 mm
95 mm
115 mm
95 mm
75 mm
MSTM/KW/0408