Descriptions 10-17
Centrifuge 5804/5804 R and 5810/5810 R
10 Descriptions
Notes for centrifuge operation
With the exception of Short spin and the programming of fixed programs, all machine functions can be
operated via the serial interface. The fixed programs of the centrifuge cannot be used, although this
function may be simulated on a computer by transferring fixed data records.
The parameters 10, 11, 12, 13, 15, 16, 17 may be modified during a centrifugation run. Unlike operation
using the keypad, the run time (Par. 12) cannot be calculated using the time which has already expired.
The value which is transferred is valid as the new nominal value.
Par. 12 =0 signifies continuous operation.
For technical reasons, it is not possible to switch during the braking phase from free deceleration
(Par. 17 =0) to a braking level (Par.17= 1...9).
MSTM/KW/0408