Error lists/Troubleshooting 3-1
Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
Error list 58XX
Display Cause Remedy
Error 1 Error in measurement system for
rotational speed.
No rotor recognized after 10
measurements.
No starting magnet (SM+L1)
recognized when calculating the
rotational speed.
Rational speed for the measurement is
not reached, abortion after 12 sec.
The motor or the frequency converter
could be defective.
The time for one rotation is < 4 ms
(errors).
Device brakes.
–After repair:
False connection of cable U2/V2/W2
from motor to converter.
Check the rotational system on the
rotor.
Service program 8.
Check/replace motor cap incl. speed
sensor, check connections.
Replace "Control" PCB,
check connections at plug X10.
Service program 9
converter
motor run.
Customer service program 9
Check connections.