3 Error lists / Troubleshooting
3-2 Error lists/Troubleshooting
Error 2 Imbalance sensor is defective.
The idle value is outside of the limits or
the measuring value is absolutely incor-
rect. Error occurs only if an incorrect idle
value is applied for at least 3 sec.
Device does not start or brake.
The old Imbalance-recognition PCB is
still in use (cable colors at plug X 10:
white/black/red/blue).
Is the customer´s rotor damaged or
loaded asymmetrically?
Is the centrifuge placed on a
horizontal, stable and resonance-free
work surface?
Was the centrifuge moved and just put
into operation? If so, let the centrifuge
run at maximum speed.
Is the sensor idle value stable and
within the tolerance (670 ±100)?
Adjust imbalance PCB and carry out
imbalance adjustment. Replace
imbalance PCB in case of no success.
Was the imbalance PCB adjusted
correctly after being replaced or after
the motor being replaced?
Following replacement of the motor,
the PCB Imbalance or the PCB
Control, was a fresh imbalance
calibration performed?
Was the correct rotor with the correct
weighs used for an imbalance
adjustment?
Was the plug reconnected correctly to
the PCB after a repair?
Check plug X10.
–Use new Imbalance-recognition PCB
from -02 onwards (cable colors at plug
X 10: White/brown/green/yellow).
Display Cause Remedy