Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
3-14 Error lists/Troubleshooting
Overload Drive overloaded. – See Error 6.
– Mains voltage too low.
CLEAn
ro
W-Error 32
200 runs
Error appears after run 200/201/202
(from software 2.0 onwards).
Clean rotor, rotor cone, motor
cone and chamber,
grease rotor axle slightly!
With the swing-bucket rotor,
lightly grease the pivots and
grooves of the bucket with
pivot grease.
– Open lid.
Standby
off
Centrifuge not used for 6 hours
(from software 2.0 onwards).
Open lid and close again.
Display Cause
MSTM/0408