Error lists/Troubleshooting 3-21
Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
Service programs 58XX
General description
The service programs run independently from the normal operating modes.
You enter the program selection by simultaneously pressing the PLUS-/MINUS key and switching the
centrifuge on.
Display:
xx Service program number
The required program number can be set by actuating the PLUS or MINUS key.
You can open the lid with the OPEN button.
All programs are started with the START button.
Indicated errors can be erased with the START button.
If "error 36" is displayed immediately after calling up the service programs the EEPROM must be
initialized (see section Initializing).
The service mode can only be quit by switching off the device.