Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
3-30 Error lists/Troubleshooting
Determination of the acceleration and braking times (for factory checks)
START Program 11
completely equip the rotor
(the measured rotational speed is calculated by the program after the rotor recognition = 97 % of
the max. rotational speed of the rotor in use)
close the lid
START run
Display of the acceleration time in seconds (at 97 % of the rotational speed nominal value)
STOP start of the measurement of the braking time
Display braking time
STOP return to the program selection
Overspeed cut-off test (Test with Rotor)
START Program 12
Close the lid
The software sets a rotational speed above the switching-off limit. If the device operates correctly, it
switches itself off when reaching the switching-off rotational speed and stops without braking.
If the device is not switched off, the run must be stopped immediately by hand!
The device must be switched off!
On switching on again the centrifuge brakes and is then again in the normal operational mode.
Overspeed cut-off:
The overspeed cut-off is about 300 1/min above the nominal speed of the rotor in use.
30 0 2000 11
HL x x x
BL x x x
MSTM/0408