Error lists/Troubleshooting 3-29
Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
Overview of rotors required and weights: G1/G2/G3
Fixed-angle rotors (W)
Centrifuges: 5804 / 5804 R / 5810 / 5810 R
Swing-bucket rotors (A)
Centrifuges: 5804
Centrifuges: 5804 R
Centrifuges: 5810 / 5810 R
Rotor designation G1 G2 G3
W F-34-6-38 6g 10g 12g
W F-45-30-11 6g 10g 15g
W FA-45-30-11 6g 10g 15g
W T-60-11 7.5g 15g 22.5g
W F-45-48-PCR Without PCR strips
a PCR strip filled
with water
Not required
Rotor designation G1 G2 G3
A A-4-44 16g 22g 25g
A A-2-DWP 16g 22g 25g
A A-2-MTP 16g 22g 25g
Rotor designation G1 G2 G3
A A-4-44 12g 18g 25g
A A-2-DWP 12g 18g 25g
A A-2-MTP 12g 18g 25g
Rotor designation G1 G2 G3
A A-4-62 16g 22g 30g
A A-4-44 16g 22g 30g
A A-2-DWP 16g 22g 30g
A A-4-62 MTP 16g 22g 30g
A A-2-MTP 16g 22g 30g
A A-4-81 16g 22g 30g
A A-4-81 MTP 16g 22g 30g
A A-4-81 Flex 16g 22g 30g