25
Installation
Concentrator plus/Vacufuge® plus
English (EN)
4Installation
4.1 Selecting the location
Select the device location according to the following criteria:
•Suitable power connection as per the name plate (230 V/120 V/100 V).
•Stable, horizontal and resonance-free lab bench.
•A well ventilated environment which is protected from direct sunlight to prevent the device from heating
up more.
•Ambient temperature during operation: 15 to 35 °C.
•At altitudes 1000 m above MSL (risk of insufficient cooling air supply), measures in accordance with
DIN EN 60034-1; VDE 0530-1 are required.
4.2 Preparing installation
4.2.1 Opening the package
Carry out the following steps in the order described:
1. Cut the adhesive tape.
2. Open the four cardboard flaps.
3. Remove accessories and rotors.
4. Remove the cardboard cover.
5. Cut the strap retainers.
NOTICE! If an error occurs, the objects in the immediate proximity of the device will be
damaged.
In accordance with the recommendations of EN 61010-2-020, leave a safety clearance of
30 cm around the device during operation.
Please remove all materials and objects from this area.
NOTICE! Damage from overheating.
Do not install the device near to any heat sources (e.g., heating, drying cabinet).
Do not expose the device to direct sunlight.
Ensure unobstructed air circulation. Keep free a clearance of at least 30 cm around all
ventilation grilles.
The weight of the complete system is 31.5 kg. The weight of the basic device is 16.5 kg.
•The device has to be transported and installed by two people.
•Only transport the device in its original packaging.
•Use a transport aid for longer distances (e.g., a hand truck).
•Retain the packaging and the transport protection device for subsequent transport or
storage. See also the instructions relating to transport (see p. 53).