Operation
Concentrator plus/Vacufuge® plus
English (EN)
40
5.3.6 Closing the lid of the device
1. Check that the rotor is correctly positioned.
2. Close the lid of the device.
5.4 Concentration
Each of the applications described here must be preceded by the preparation described above (see
Preparation for concentration on p. 36).
WARNING! Device lid can crush. Keep hands clear.
Do not reach between the device and lid when opening or closing the device lid.
WARNING! Risk of injury from improperly attached rotors.
Only operate the device if the rotor has been mounted properly.
If there are any unusual noises when the device is started up, immediately press the start/
stop key to stop concentrating.
CAUTION! Risk from incorrectly loaded rotors and damaged/overloaded tubes!
Before commencing concentration, follow the safety instructions relating to risks from
asymmetrically loaded and/or overloaded rotors and from overloaded and/or damaged
tubes (see Warnings for intended use on p. 18).
NOTICE! Overflowing of the emission condenser.
The maximum filling volume of the emission condenser is 280 mL.
1. Check before each run whether the volume of the emission condenser is sufficient for your
application.
2. Empty the emission condenser if necessary .
Unusual noise may occur when braking heavy rotors. This is construction-related and does
not indicate a problem.