55
Technical data
Concentrator plus/Vacufuge® plus
English (EN)
9Technical data
9.1 Power supply
9.2 Ambient conditions
9.3 Weight/dimensions
5305 Basic device 5305 Complete system
Mains connection: 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
Current consumption: 2.1 A (230 V)
6.0 A (120 V)
5.2 A (100 V)
1.7 A (230 V)
5.6 A (120 V)
3.8 A (100 V)
Power consumption: max. 500 W (230 V)
max. 550 W (120 V)
max. 520 W (100 V)
max. 350 W (230 V)
max. 400 W (120 V)
max. 380 W (100 V)
Overvoltage category: II
Fuses: 4.0 AT (230 V)
6.3 AT (120 V/100 V)
Environment: Only for use indoors.
Ambient temperature: 15 to 35 °C
Max. relative humidity: 75 %, non-condensing humidity
Atmospheric pressure: 79,5 kPa – 106 kPa
Degree of pollution 2
Basic device Complete system with connection for
external device
Dimensions: Width: 330 mm (12.6 in.)
Depth: 372 mm (14.5 in.)
Height: 231 mm (9.1 in.)
Width: 330 mm (12.6 in.)
Depth: 579 mm (20.9 in.)
Height: 290 mm (11.7 in.)
Weight without rotor: 16.5 kg (37.5 lbs.) 31.5 kg (68.3 lbs.)
Noise level: 50 dB(A) 50 dB(A)