EasyManua.ls Logo

Eppendorf Concentrator plus - 9.4 Diaphragm Vacuum Pump of the Complete System; 9.5 Application Parameters

Default Icon
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data
Concentrator plus/Vacufuge® plus
English (EN)
56
9.4 Diaphragm vacuum pump of the complete system
9.5 Application parameters
Max. current: 1.4/1.6 A (230 V, 50/60 Hz)
3.0 A (120 V, 50 Hz)
3.6/3.8 A (100 V, 50/60 Hz)
Max. power: 180 W (230/120/100 V)
Motor protection: Thermal winding protection
Protection class according to IEC 529: IP 54
Suction capacity (in accordance with DIN EN
60529; VDE 0470-1):
1.9/2.1 m
3
/h (50/60 Hz)
Achievable ultimate pressure (absolute): <20 mbar
Max. permissible pressure at the outlet (absolute): 2 bar
Max. differential pressure between inlet and outlet: 1 bar
Maximum speed at 50/60 Hz: 1,500/1,800 rpm
Radio interference protection class: B
Materials of the surfaces that are in contact with the medium
Inner part of housing cover: PTFE, with carbon reinforcement
Head cover, diaphragm clamping disk: ETFE, with carbon reinforcement
Valve: FFKM
Diaphragm: PTFE
Inlet: ETFE
Outlet: ETFE
Screw connection: ETFE
Hose: PTFE
Run time: 1 min to 9:59 h, unlimited (oo),
adjustable in 1 min. increments.
Rotational speed: 1,400 rpm, not adjustable
Max. relative centrifugal force
(RZB or rcf):
248 x g, not adjustable
Max. load: 144 tubes/2 microplates (0.2 to 50 mL)
Permissible density of the material for centrifuging
(at max. g-force/speed and max. load):
1.2 g/mL
Emission condenser volume: 280 mL

Related product manuals