5 Description of menus
5.5 Lid
The "Lid" menu in the main menu can be used to determine the temperature of the heated lid and to switch
the heated lid on or off (e.g. to preheat the lid).
LidTemp: Nominal/actual temperature.To enter the temperature for the heated lid.
The lid heating switches off when a temperature value of 0
°
C is entered.
Heater is On: Lid heating is on.
Heater is Off: Lid heating is off.
When sensitive PCRs are carried out, it may be important to delay the start until the lid temperature has
prevented condensation. This can be achieved by programming the "WAIT" setting, which means that the
program does not start until the lid reaches the specified temperature.
In addition, the lid can also be preheated. After the device has been switched on, the "Lid" menu is selected,
the desired lid temperature is entered and "Heater is On" is selected by pressing the key.
When a program starts, it is the program settings which have priority.
When a program is running, the lid heating cannot be switched on or off.
Main-Menu
Start
FILES . .
OPTIONS . .
Lid
Incubate
STANDARD
– –/22.0°
Lid Temp
LidTemp: 105/55°C
Heater is Off