25
7 Programming
7.1 Command description
A program contains two basic commands for controlling the temperature of the thermoblock and of the heated
lid as well as six different commands for programming.
A program can contain up to 40 program lines. The commands may be repeated as desired.
Digits or letters may be entered or selections are made using the key.
When digits are entered, the message "Value out of range" appears when values are entered which are not
within the given specifications. The permitted limit value is then entered automatically.
7.1.1 CNTRL
With this command is determined the type of temperature control for the block.
BLOCK
The temperature is measured on the thermoblock and the nominal temperature is then set.
TUBE
The software adapts the temperature of the thermoblock to the temperature of the tubes and the
sample quantity. For this reason, the tube type and the filling volume must be entered
immediately after the start of the program. No further entries are necessary when
in situ
is
selected.
Tube type 0.2 ml or plate: 5 to 50
µ
l filling volume
0.5 ml or 0.5 ml thin: 5 to 100
µ
l filling volume
FILES... Edit UNNAMED
Edit CNTRL BLOCK
Load LID=0°
Standard NOWAIT AUTO
New
Delete
UNNAMED
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
1 T=❋❋❋❋° ❋❋:❋❋:❋❋
+0.0° +0:00
R=3.0°/s +0.0°/s
G=0.0°
2 Hold ❋❋❋❋° ENTER
3 PAUSE PRESS ENTER
4 GOTO
❋❋❋
REP
❋❋❋
5 SOUND
❋❋
6 LINK
❋❋❋❋❋❋❋❋
end
– –/22.0° 10:22:06
7 Programming