Technical data
Mastercycler
®
nexus
English (EN)
100
Mastercycler nexus flat, Mastercycler nexus flat eco
Sample capacity Flat block, aluminum
4 slides
Block temperature control range 4 °C to 99 °C
Ramp rate
(measured at the block)
max. 3°C/s
Block temperature homogeneity
(with the gradient function switched off)
at 35°C
at 90°C
≤ ±0.3°C
≤ ±0.4 °C
Control accuracy
(with the gradient function switched off)
±0.2°C
Gradient range max. 20°C
Gradient temperature control range 30 °C to 99 °C
Heated lid temperature control range 37 °C to 110 °C
Mastercycler nexus GX2, Mastercycler nexus GX2e
Sample capacity Universal block, aluminum
96 × 0.1 mL or 0.2 mL PCR tubes
64 × 0.5 mL PCR tubes
one 64-well PCR plate, one 32-well PCR plate (unskirted,
semi-skirted, skirted; standard and low-profile)
Block temperature control range 4 °C to 99 °C
Ramp rate
(measured at the block)
max. 3°C/s
Block temperature homogeneity
(with the gradient function switched off)
at 35°C
at 90°C
≤ ±0.3°C
≤ ±0.4 °C
Control accuracy
(with the gradient function switched off)
±0.2°C
Thermoblock (64 wells)
Gradient range max. 12°C
Gradient temperature control range 30 °C to 99 °C
Heated lid temperature control range 37 °C to 110 °C
Thermoblock (32 wells) no gradient function
Gradient range –
Gradient temperature control range –
Heated lid temperature control range 37 °C to 110 °C