Technical data
Mastercycler
®
nexus
English (EN)
101
Mastercycler nexus X2, Mastercycler nexus X2e
Sample capacity Universal block, aluminum
96 × 0.1 mL or 0.2 mL PCR tubes
64 × 0.5 mL PCR tubes
one 64-well PCR plate, one 32-well PCR plate (unskirted,
semi-skirted, skirted; standard and low-profile)
Block temperature control range 4 °C to 99 °C
Ramp rate
(measured at the block)
max. 3°C/s
Block temperature homogeneity
at 35°C
at 90°C
≤ ±0.3°C
≤ ±0.4 °C
Control accuracy ±0.2°C
Gradient range –
Gradient temperature control range –
Heated lid temperature control range 37 °C to 110 °C