4-10
4.3.1 Unit-related output error
The ICC is equipped with automatic monitoring of the HF output
parameters, which monitors deviations in the actual value from
the set value for the currently set HF output parameters and
produces a warning signal and/or switches off the HF generator
if the deviation is so large than the required quality of the
respective effect (cutting or coagulation) is no longer ensured. In
case of deviations or absence of the required effect, display of a
unit-related output error allows the operator to immediately check
whether the defect is caused by the unit or not.
Deviations in the HF output parameters from the currently set HF output parameters on the ICC
can only result from loads which have too low a resistance, e.g. coagulation electrodes that are
too large, a short circuit between the active and neutral electrode, or due to an error in the unit.
4.3. Description of the safety features
The ICC 300 is equipped with the following safety features:
4.3.1 OUTPUT ERROR = Automatic monitoring of unit-related output error
4.3.2 TIME LIMIT = Automatic monitoring of the time limit
4.3.3 NESSY = Neutral Electrode Safety System
4.3.4 Protection against operating error