8-3
Error list
Error
No.
Error How to proceed
0 No error
1 No HF output voltage Notify the Technical Service
2 HF output voltage too high Notify the Technical Service
3 No HF output voltage Notify the Technical Service
4 HF output voltage too high Notify the Technical Service
6 Activation error Notify the Technical Service
7 Activation error Notify the Technical Service
9 Time limit exceeded Heed maximum time limit
10 Erroneous setting of the AUTO CUT
function field during activation
Before activation of a cutting mode, the AUTO CUT
function field must be completely set
11 Erroneous setting of the AUTO COAG
function field during activation
Before activation, the AUTO COAG function field
must be completely set
12 Erroneous setting of the AUTO
BIPOLAR function field during
activation
Before activation, the AUTO BIPOLAR function field
must be completely set
13 The contact surface between the neutral
electrode and the patient is too small or
the neutral electrode was not connected to
the unit.
This error is recognized and reported in
the Standby mode or during activation of
the unit.
Only for NESSY Version 2
Error in the connection between the
neutral electrode and unit
Only for NESSY Version 3
Short between the two contact surfaces of
a divided neutral electrode or in the cable
to the neutral electrode
Use a sufficiently large neutral electrode and apply the
entire surface.
Check the connection of the neutral electrode to the
unit.
Check the connection between the neutral electrode
and unit. Probably a faulty cable or faulty plug
Eliminate short or exchange cable
19 The contact surface between the neutral
electrode and the patient was too small
Use a sufficiently large neutral electrode and apply the
entire surface.
20 The neutral electrode was applied in the
incorrect direction
See Chapter 4.3.3
NESSY
21 The neutral electrode was applied in the
incorrect direction
See Chapter 4.3.3
NESSY
22 When switching on the power switch, the
yellow pedal on the footswitch or the
yellow key on the electrode handle was
already pressed
Check whether there was an operator error or whether
the yellow pedal on the footswitch or the yellow key
on the electrode handle is defective
23 When switching on the power switch, the
blue pedal on the footswitch or the blue
key on the electrode handle was already
pressed for CUT/COAG
Check whether there was an operating error or
whether the blue key on the electrode handle or the
cable to the electrode handle is defective
24 When switching on the power switch, the
yellow pedal on the footswitch or the
yellow key on the electrode handle was
already pressed for CUT/COAG
Check whether there was an operating error or
whether the yellow pedal on the footswitch or the
yellow key on the electrode handle is defective