Weighing Parameters
With pAr on display press key. Weighing parameters are edited one at a time. After that each press of
the key moves to the next parameter. Edit the values using the and keys.
Capacity 1 (Max 1)
cAP_1
In single range instruments, this is the maximum capacity.
In multi (two) range instruments, this is the maximum capacity of the first range
(smaller range).
Capacity 2 (Max 2)
cAP_2
In single range instruments, this value have to be entered as 0 (zero)
In multi (two) range instruments, this is the maximum capacity of the second range
(greater range).
Scale interval (e)
E 2
This value is the difference between two consecutive indicated values (e).
In multi range instruments, this is the interval of the smaller range (e1). The device
selects e2 automatically.
Accepted values: 1,2,5,10,20,50
Decimal point
. 3
This value determines the place of the decimal point while indicating weight values.
Accepted values: 0,1,2,3
Stability range
Stb_r 1
The range for displayed weight values, for determination of the stability.
Accepted values: 0 .. 5 intervals (Note: A value of zero means 0.5 interval)
Stability delay
Stb_d 1
The time period for displayed weight values, for determination of the stability.
Accepted values: 1 .. 5 seconds
Filter
FıL 3
This value determines the interpretation of the changes in weight. A lower value
causes fast reactions to weight changes and is suitable in relatively stable conditions.
A higher value causes slow reactions but the vibrational flickers are eliminated.
Accepted values: 1 (most sensitive, fast reaction) .. 5 (most coarse, slow reaction)
Zero at power-on
ZP 1
This setting determines if the indication is set to zero automatically at power on.
(initial zero setting)
Accepted values: 0 (off) , 1 (on)
9