Adjustment
Press key while AdJuSt is displayed. There are 3 options for adjustment:
Dead load
adjustment
This option may be used when only dead load (zero point) adjustment is needed.
Zr_cAL
While this message is displayed, unload the platform and press key. The display
will become blank until the device calculates and saves the adjustment data. After
that, the device exits the menu and returns to the weight display.
Span adjustment
This option may be used when a dead load adjustment is not needed, but a span
adjustment is needed.
Zr_cAL
While this message is displayed, press key. The dead load adjustment is
skipped.
010000
Enter (edit) the value of the weight which will be used for adjustment and press
key.
10000
The value you entered will be displayed constantly. Load the platform with the
adjustment weight you entered and press key when there is no motion. The
display will become blank until the device calculates and saves the adjustment data.
After that, the device exits the menu and returns to the weight display.
Full adjustment
This option is performing the dead load adjustment and the span adjustment
consecutively.
Zr_cAL
While this message is displayed, unload the platform and press key when there
is no motion.
010000
Enter (edit) the value of the weight which will be used for adjustment and press
key.
10000
The value you entered will be displayed constantly. Load the platform with the
adjustment weight you entered and press key when there is no motion. The
display will become blank until the device calculates and saves the adjustment data.
After that, the device exits the menu and returns to the weight display.
10