Escort ELC-3133A - Capacitance (Parallel mode) Accuracy

Escort ELC-3133A
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Capacitance (Parallel mode)
Test Frequency: 100 / 120 Hz
Accuracy Range Maximum
Display
Capacitance DF
Spec. Note
10mF 19.99mF
*N4
2.5%+5
(DF<0.1)
5%+100/Cx+5
(DF<0.1)
After short
cal.
1000
μ
F 1999.9
μ
F
*N5
0.6%+5
(DF<0.1)
1%+100/Cx+5
(DF<0.1)
After short
cal.
200
μ
F 199.99
μ
F
0.4%+3
DF<0.5
0.4%+100/Cx+5
(DF<0.5)
-
20
μ
F 19.999
μ
F
0.4%+3
(DF<0.5)
0.4%+100/Cx+5
(DF<0.5)
-
2000nF 1999.9nF 0.4%+3
(DF<0.5)
0.4%+100/Cx+5
(DF<0.5)
-
200nF 199.99nF 0.4%+5
(DF<0.5)
0.4%+100/Cx+5
(DF<0.5)
After open
cal.
20nF 19.999nF 0.6%+5
(DF<0.1)
1%+100/Cx+5
(DF<0.1)
After open
cal.
Test Frequency: 1 KHz
Accuracy
Range Maximum
Display
Capacitance DF
Spec. Note
1mF 1.999mF
*N4
2.5%+5
(DF<0.1)
5%+100/Cx+5
(DF<0.1)
After short
cal.
200
μ
F 199.99
μ
F
0.6%+5
(DF<0.1)
1.2%+100/Cx+5
(DF<0.1)
After short
cal.
20
μ
F 19.999
μ
F
0.4%+3
(DF<0.5)
0.4%+100/Cx+5
(DF<0.5)
-
2000nF 1999.9nF 0.4%+3
(DF<0.5)
0.4%+100/Cx+5
(DF<0.5)
-
200nF 199.99nF 0.4%+3
(DF<0.5)
0.4%+100/Cx+5
(DF<0.5)
-
20nF 19.999nF 0.4%+5
(DF<0.5)
0.4%+100/Cx+5
(DF<0.5)
After open
cal.
2000pF 1999.9pF 0.6%+5
(DF<0.1)
1.0%+100/Cx+5
(DF<0.1)
After open
cal.