EasyManua.ls Logo

ESI 250-DA - Understanding and Resolving Sliding Balance

ESI 250-DA
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2,3.7
SLIDING
BALANCE
WHAT
IT
IS
CHANGING
THE
L
DIAL
CHANGING
THE
Q
DIAL
When
lossy
inductors
are
measured
the nulling
problem
known
as slid-
ing
balance
will
occur
if
the
usual
procedure
of
adjusting
first
one
con-
trol
then
the
other
is
followed.
As shown in the
above
drawings,
neither
the
Q
dial
nor
the L
dial
will
move
the
bridge
vector
toward
a null
directly
or
quickly.
When
Q
is
high
there
is
no sliding
balance
because
the
effects
of
Q
and
L
are
perpendicular to
each
other and
nulling is
easy.
WHAT
TO
DO
ABOUT
IT
WHEN
USING
ESI
250-DA
BRIDGE
This
problem
is
not
too
serious
if the
value
of
Q
is
greater
than
one. For
lower
values
of
Q,
an
external
D-Q
rheostat
may be
necessary.
For
low
Q,
the
L
series
circuit
should
be
used.
The
following
approach
should
be
used
if
balancing
is
difficult.
By
a
somewhat
different
nulling
technique
rapid
accurate
low
Q
measure-
ments can
be
accomplished.
First
set
the
Q
dial
to
its
maximum
position,
then
set L
for a
detector
minimum.
With
Q
set
high,
L
will
move
the
impedance
vector in
the L
direction
only.
The
bridge
null
detector
signal
is
almost
exactly
pro-
portional
to
the
difference
between
the
unknown
impedance
and
the im-
pedance
indicated
by
the
bridge
dial.
The
correct
L
setting
will
bring
the
impedances
closer
together
and
give
minimum
signal.
Next
set the
Q
dial
for
a
minimum
on
the
detector.
This
should
give
a
good start
on
finding
L
and
Q.
Move
L
a
small
amount
then set
Q
for a
minimum
and
observe
whether
this
minimum
is
better
or
worse
than
the
previous
one.
If
it
is
better
move
L
another
small
amount
in the
same
direction.
If it
is
worse try
the
other
direction.
Continue
moving
L
in
small
steps
in
one
^i^
2
^ection
only
while
setting
Q
for
minimum
detector
signal
until the
detector
signal
reaches
a
true
null.
Moving
L
further
will
make the
null
worse
again.
The
null
found
this
way
is
correct.
A
little
practice
will
allow
you
to
make
these
measurements
rapidly
even
when
Q
is in
the
order
of
0,1.
When
almost
nulled
it
is
often
easier to
move
the
Q
dial
"one
wire
at a
time"
while
null
seeking
with
the
center
L dial.
Be
careful with
the
L
series
circuit.
The
bridge
will
also
indicate
a null at
a setting
of
Q
=
L
=
O
because
at
this
setting
the
generator
is
shorted.
|
e
|
s
]
i
]
250-DA
7/59

Related product manuals