EasyManua.ls Logo

ESI 250-DA - Understanding and Resolving Sliding Balance in Capacitance Measurements

ESI 250-DA
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.4.7
SLIDING
BALANCE
WHAT
IT
IS
u
FIXED
unknown
d
setting
impedance
J;
CHANGING
A
X
RANGE
OF
BRIDGE
C
SETTING
I'
impedance
R
CHANGING
THE
C
DIAL
CHANGING
THE
D
DIAL
When
lossy
capacitors
are
measured
the
nulling problem
known as
slid-
ing
balance
will
occur
if
the
usual
procedure
of
adjusting
first one con-
trol
then
the
other
is
followed.
As shown in
the above
drawings,
neither
the
D dial
nor
the
C
dial
will
move
the
bridge
vector
toward
a null
directly
or
quickly.
When D
is
low there
is
no sliding balance
because
the
effects
of D
and C
are
perpendicular
to each
other and nulling
is easy.
WHAT
TO
DO
ABOUT IT
WHEN
USING
ESI
250- DA BRIDGE
This
problem
is
not
too
serious
if the
value
of
D is less than
one. For
higher
values
of D an
external
D-Q
rheostat
may
be
necessary. For
high
D,
the
C
parallel
circuit
should
be
used,
(see
2.4.8)
The
following
approach
should
be
used
if
balancing
is
difficult.
By a
somewhat
different
nulling
technique
rapid
accurate
high D
measure-
ments
can
be
accomplished.
First
set
the
D dial
to
its
minimum
position,
then set C
for a detector
minimum.
With
D
set
low,
C
will
move
the
impedance
vector
in the C
direction
only.
The
bridge
null
detector
signal
is
almost
exactly
propor-
tional to
the
difference
between
the
unknown
impedance and
the impedance
indicated
by the
bridge
dial.
The
correct
C
setting
will bring
the
imped-
ances
closer
together
and
give
minimum
signal.
Next
set the D
dial
for a
minimum
on
the
detector.
This
should
give a
good start
on finding
C
and
D.
Move C
a
small
amount
then
set D for
a
minimum
and
observe
whether
this
minimum
is
better
or
worse
than
the
previous
one. If it
is
better
move C
another
small
amount
in
the
same
direction.
If it
is worse
try
the
other
direction.
Continue
moving C
in
small
steps in
one
direction
only
while
setting
D
for
minimum
detector
signal until
the
detector
signal
reaches
a
true
null.
Moving
C
further
will
make
the
null
worse
again.
The
null
found
this
way
is
correct.
A
little
practice
will
allow
you to
make
these
measurements
rapidly
even
when D
is
in the
order of 10,
When al-
most
nulled
it
is
often
easier to
move
the
D
dial
"one wire at
a time"
while
null
seeking
with
the
center
C
dial.
Be
careful with the C
parallel circuit.
The
bridge
will
also
indicate
a
null at
a
setting
of
C
=
i
=
O because at
this
setting
the
generator
is
shorted.
felHTI
250-DA
7/59

Related product manuals