6.2.9
CHECKING
THE
MAIN
DIAL
DEKASTAT°
Set
the
controls
as
follows:
CONTROL
SETTING
DETECTOR
switch
AC
OR
EXT DET
GENERATOR
switch
EXT
GEN
CIRCUIT
SELECTOR
switch
R X 1
L-R
HIGH
TERMINAL;
EXTERNAL
D-Q
TERMINAL
MARKED
G.
(D-Q
shorting
bar
must
be
in
position.)
Connect
ohmmeter test
leads
The
ohmmeter
reading
in
kilohms
should agree
with
the
MAIN
DIAL
setting
.
6.2.10
CHECKING
THE
D-Q
RHEOSTATS
6.
2.
10.
1
To
check
set
the
controls
as
follows:
CONTROL
SETTING
DETECTOR
switch
AC
OR
EXT
DET
GENERATOR
switch
EXT
GEN
CIRCUIT
SELECTOR
switch
L
-
SERIES
MULTIPLIER
switch
100 h
D-Q
DIAL
10
C
HIGH
TERMINAL,
EXT
D-Q
TERMINAL
MARKED
G.
(short-
ing
bar
must
be in
position)
Connect
ohmmeter
leads
The
ohmmeter
should
read
about
16 K.
lelsi i
\ 250
-DA 6/5
9