Chapter 3 Commands
■To Monitor Constant Mode Pattern No.
Table 3.11 Constant mode pattern No. monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the current pattern
No. being used in the constant mode.
<
Example command
>
“1,CPAT?”
<
Response format
>
“pattern No.”
1 to 16 CPAT?
−
<
Example response
>
“1”
■To Monitor Constant Mode Parameters
Table 3.12 Constant mode parameters monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the target
temperature, humidity and time set for the constant mode.
<
Example command
>
“1,SET?”
<
Response format
>
“target temperature,target humidity,target time”
1 to 16 SET?
−
<
Example response
>
“105.0,100,100.0”
■To Monitor Constant Mode Time Information
Table 3.13 Constant mode time information monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the target time set
for the constant mode and the time remaining till mode end.
<
Example command
>
“1,TIME?”
<
Response format
>
“target time,remaining time”
1 to 16 TIME?
−
<
Example response
>
“100.0,50.0”
26