10 P-300 Web Controller Configuration and Troubleshooting Manual
The first step in the web controller configurations process is to find out the details for all of the various
configurations and record them for use during the config procedure. Fill in the following data chart for the
configurations of your test chamber and controller.
Chamber Model Details
Item Number
Configuration
Parameter
P-300 Screen to use for
Verification Actual Data
1
Test chamber has humidity
(yes or no)
Constant setup screen
2
If the chamber has humidity,
does it have “low humidity”
or “ultra low humidity”
options
This data is not on the P-300
screens, verify actual with
chamber specification or
drawings
3
Test chamber has product
temperature option (yes or
no)
Constant setup screen
4
Chamber type or model (EP,
EGN, EN, EWS, EWP)
See list below for details
This data is not on the P-300
screens, verify actual with
chamber specification or
drawings
Model Name
Model “code”
(used by web server config screen)
Platinous H EP
Platinum EN
Global N-series EGN
Solid Walkin EWS
Panelized Walkin EWP