P-300 Web Controller Configuration and Troubleshooting Manual 13
2.2 Additional Web Controller Configuration Functions
2.2.1 Custom Temp-Humidity Performance Graphic Files
Custom Graph
1. Click on the “Browse” button in order to use a “non-standard” temperature humidity performance graph
for the image that is used on the “Constant Mode Setup” screen.
2. Navigate to the appropriate graphic file, and then click the open box. Click the “Upload File” button to
upload the image to the web server you are connected to. The file will upload and a confirmation dialog
box will appear to confirm completion of the upload.
3. Click the “OK” button. After the confirmation dialog box disappears, you will see that “999_custom.gif” is
selected as the new “Graph file” in the manual edit config settings area in the top section of the screen.
4. Click the “Save” button, and then the “Reboot chamber controller” button to complete your new graphic
installation process.
2.2.2 Uploading and Downloading Configuration Parameter Files
Download Configuration
1. Click on the “Download current configuration” button to download the configuration parameters from the
web controller you are connected to and storing the file on a server or other hard drive.
This feature can be useful if you have a number of chambers that all need a custom configuration. You can
create the configuration for one of the web controllers, download the file, and then upload it to all of the
other web controllers that require the same custom configuration. This feature can also be useful for
historical archiving purposes.
Upload Configuration
1. Click on the “Browse” button in order to select a web controller configuration parameters file for upload
to the currently connected web controller.
2. Navigate to the appropriate configuration file, and then click the open box. Click the “Upload File” button
to upload the configuration to the web server you are connected to. The file will upload and a confirma-
tion dialog box will appear to confirm completion of the upload.
3. Click the “OK” button. After the confirmation dialog box disappears, you will see the individual controls
in the upper “Edit Config” section of the screen update with the new parameters.
4. Click the “Save” button, and then the “Reboot chamber controller” button (at the bottom of the page) to
complete the process.