Platinous User’s Manual 91
Troubleshooting
Displayed Alarm
Code/Channel
Code (option)
Trouble Cause Remedial Action
OUT-OF-RANGE
(AI-CH5 ~ CH12)
(8)
A temperature outside of the
specified range was detected
on one of the temperature
controller channels (Ai5 ~
12ch) that monitor the
refrigeration circuit. The
chamber has been stopped.
If the chamber has more than
1 refrigerator and the backup
mode is ON, only the affected
refrigerator will shut down,
while the chamber will
continue running using
another refrigerator.
Frost buildup,
compressor
trouble or a
refrigerant leak
are all possible
causes.
After testing ends, turn
control power OFF and ON
from the POWER key and
resume testing. If the same
alarm occurs again, call for
service.
HUMIDIFIER
BOIL DRY
The humidifier thermostat
opened because the
operating temperature was
too high.
Low humidity tray
water level.
Lack of water
supply or mineral
build-up in
humidity tray.
Set the main power switch
(leakage breaker) in the
OFF position and check/
adjust water level. Check
water supply. Clean
humidity tray. Reactivate
the system from the main
power switch followed by
the
POWER key, and resume
testing. If the same alarm
occurs again, call for
service.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ESPEC Platinous ESL-4C Series and is the answer not in the manual?

ESPEC Platinous ESL-4C Series Specifications

General IconGeneral
BrandESPEC
ModelPlatinous ESL-4C Series
CategoryClimate chamber
LanguageEnglish

Related product manuals