Bench-top type temperature(humidity) chamber Installation guide
8 ESPEC CORP.
How to install
Notice
Do not move the chamber with specimens inside.
The specimens may turn over or fall off the shelves.
Install the chamber only on a flat, level floor.
An unlevel chamber can possibly cause water to backflow without smooth drainage
from the dew tray, mistakenly trigger alarms, and destabilize humidity control.
Leveling feet are found under each of the four corners of the chamber base.
< Procedure >
1) Turn the adjustable foot until at a height that the chamber sits level.
2) Place a level on the center of the test area floor, and check the level.
Levelling feet
To raise To lower