7-19 Cannot measure PD (cannot measure PD correctly)
Main causes when it cannot measure PD (cannot measure PD correctly:
・ Failure of PD meter ASSY
・ Failure of control board
Replace the PD meter ASSY.
Is the performance of PD
meter ASSY normal?
Check the measurement start switch.
【Refer to 7-11.】
Is the measurement start switch normal?
(Is the refractivity measurable?)
Replace the control board.
Is the peripheral relay
harness connected correctly?
Connect the peripheral
relay harness correctly.