III. Specification
1 Model of Series
Model
Jaw
Size
(mm)
Range of
Resistance
(Ω)
Range of
Current
(A)
Storage
Function
Alarm
Function
ETCR2000+ 65×32
0.01--1200 -- 99 Units
ETCR2000A+
65×32
0.01--200 -- 99 Units
ETCR2000C+
65×32
0.01--1200 0.0--20.0 99 Units
ETCR2100+ φ32 0.01--1200 -- 99 Units
ETCR2100A+
φ32 0.01--200 -- 99 Units
ETCR2100C+
φ32 0.01--1200 0.0--20.0 99 Units
Note: √” means available.
2. Ranges and Accuracy of Measurement
Mode Range Resolution
Accuracy
Resistance
0.010Ω-0.099Ω 0.001Ω ±1%0.01Ω
0.10Ω-0.99Ω 0.01Ω ±1%0.01Ω
1.0Ω-49.9Ω 0.1Ω ±1%0.1Ω
50.0Ω-99.5Ω 0.5Ω ±1.5%0.5Ω
100Ω-199Ω 1Ω ±2%1Ω
200Ω-395Ω 5Ω ±5%5Ω
400-590Ω 10Ω ±10%10Ω
600Ω-880Ω 20Ω ±20%20Ω
900Ω-1200Ω 30Ω ±25%30Ω
*Current
(True-RMS)
0.00mA -9.00mA 0.05mA ±2.5%2mA
10.0mA -99.0mA 0.1mA ±2.5%10mA
100mA -300mA 1mA ±2.5%20mA
0.30A-2.99A 0.01A ±2.5%0.1A
3.0A-9.9A 0.1A ±2.5%0.5A
10.0A-20.0A 0.1 A ±2.5%1 A
Resistance Measurement Frequency: >1KHz
Measured Current Frequency: 50Hz/60Hz
Setting Range of Resistance Alarm Critical Value: 1Ω-199Ω
*Setting Range of Current Alarm Critical Value1mA -499mA